Pulse leveling accuracy 3-41, Pulse leveling accuracy -41 – Anritsu 682XXB User Manual

Page 101

Advertising
background image

Pulse
Leveling
Accuracy

The following procedure lets you measure the pulsed
RF output level and CW output level and compare
them to verify pulse leveling accuracy.

1. On the 682XXB/683XXB:

a. Press

On/Off

(Internal Pulse Status menu) to

turn off pulse modulation.

b. Press

CW/SWEEP SELECT

to return to the

CW Menu display.

c. At the CW menu, press

Edit F1

and set F1 to

the frequency noted in the Test Record.

2. On the Oscilloscope:

a. Adjust the vertical offset to place the trace ex-

actly on the center graticule; this is used as the
CW-level reference line.

b. Use the Auto Triggering mode to continue

sweeping the display when the pulse is off.

c. Adjust the vertical sensitivity to the most sen-

sitive setting (mV/div) possible while keeping
the trace at the center graticule reference line.

682XXB/683XXB MM

3-41

PULSE DETECTOR OUTPUT WAVEFORM

Percentage
of Detected

Waveform

0%
2%

36%

86%

100%

114%

0%
10%

50%

90%

100%

114%

T

RISE

T

FALL

Corresponding

Percentage

of RF Voltage

PULSE WIDTH

Note: Waveform shown is for
a negative output detector.

0%

10%

OVER-

SHOOT

Figure 3-11.

Measurement Parameters for a Pulse Modulation Waveform

PERFORMANCE

PULSE MODULATION TESTS:

VERIFICATION

RISE TIME, FALL TIME, OVERSHOOT, AND LEVEL