3-15

PULSE MODULATION
TEST: VIDEO
FEEDTHROUGH

This pulse modulation test verifies that video feedthrough is within
specifications.

Test Setup

Connect the equipment, shown in Figure 3-12, as
follows:

1. Connect the 682XXB/683XXB rear panel

PULSE

SYNC OUT

to the Oscilloscope Trigger Input.

2. On the Oscilloscope, select 50

W input impedance.

3. Connect the 682XXB/683XXB

RF OUTPUT

to the

Oscilloscope Vertical Input through a 450 MHz
Low Pass Filter.

682XXB/683XXB MM

3-43

PERFORMANCE

PULSE MODULATION TEST:

VERIFICATION

VIDEO FEEDTHROUGH

450 MHz
Low Pass Filter
(For Video Feedthrough
Measurement)

CH. 4 Trigger

Pulse
Sync
Out

682XXB / 683XXB

CH. 1 Input

Oscilloscope

Figure 3-12.

Equipment Setup for Pulse Modulation Test: Video Feedthrough