PULSE MODULATION
TEST: RF ON/OFF
RATIO

This pulse modulation test verifies that the ratio of RF on power to RF
off power is within specifications.

Test Setup

Connect the equipment, shown in Figure 3-13, as
follows:

1. Connect the 682XXB/683XXB rear panel

10 MHZ

REF OUT

to the Spectrum Analyzer External Ref-

erence Input.

2. Connect the

RF OUTPUT

of the 682XXB/683XXB

to the Spectrum Analyzer RF Input.

Test
Procedure

The following procedure lets you determine the ratio
of RF on power to RF off power during pulse modu-
lation.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to the frequency noted on the Test Re-

cord.

d. Press

MODULATION

, then

Pulse

. The Inter-

nal Pulse Status menu is displayed.

682XXB/683XXB MM

3-45

PERFORMANCE

PULSE MODULATION TEST:

VERIFICATION

RF ON/OFF RATIO

10 MHz

REF OUT

682XXB / 683XXB

EXT REF

INPUT

RF OUT

Spectrum Analyzer

RF IN

Figure 3-13.

Equipment Setup for Pulse Modulation Test: RF On/Off Ratio