INITIAL SETUP

The 682XXB/683XXB is calibrated using an IBM compatible PC and
external test equipment. The PC must have the Windows 3.1 or Win-
dows 95 operating system installed and be equipped with a mouse.
Initial setup consists of interfacing the PC to the signal generator.

Interconnec-
tion

Using the ANRITSU P/N T1678 serial interface as-
sembly, connect the PC to the 682XXB/683XXB as
follows:

1. Connect the wide flat cable between the 682XXB/

683XXB rear panel

SERIAL I/O

connector and the

P1 connector on the T1678 serial interface PCB.

2. Connect the narrow flat cable between the P2

(TERM) connector on the T1678 serial interface
PCB and the COM1 or COM2 connector on the
PC. Use the RJ11-to-DB-9 or RJ11-to-DB-25
adapter, provided with the T1678 serial interface
assembly, to make the connection at the PC.

682XXB/683XXB MM

4-7

CALIBRATION

INITIAL SETUP

IBM-Compatible PC

To

682XXB / 683XXB

To
PC

P1

P2

P3

Term

T1678 Serial Interface

682XXB / 683XXB

COM1

or

COM2

Serial
I/O

Figure 4-1.

PC to 682XXB/683XXB Interconnection for Calibration