PRELIMINARY
CALIBRATION

This procedure provides the steps necessary to initially calibrate the
coarse loop, fine loop, frequency instruction, and internal DVM cir-
cuitry and the 100 MHz reference oscillator of the 682XXB/683XXB.

Equipment
Setup

Connect the equipment, shown in Figure 4-2, as fol-
lows:

1. Interface the PC to the 682XXB/683XXB by per-

forming the initial setup procedure, pages 4-7 to
4-12.

2. Connect the Frequency Counter to the 682XXB/

683XXB when directed to do so during the cali-
bration procedure.

NOTES

If the 682XXB/683XXB has option 19 in-
stalled, verify that the GPIB is configured
for the

Native

external interface language

before beginning instrument calibration.

Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.

682XXB/683XXB MM

4-13

PRELIMINARY

CALIBRATION

CALIBRATION

Figure 4-2.

Equipment Setup for Preliminary Calibration