4-8

SWITCHED FILTER
SHAPER CALIBRATION

This procedure provides the steps necessary to adjust the Switched
Filter Shaper Amplifier gain to produce a more constant level ampli-
fier gain with power level changes.

Equipment
Setup

Connect the equipment, shown in Figure 4-3, as fol-
lows:

1. Interface the PC to the 682XXB/683XXB by per-

forming the initial setup procedure, pages 4-7 to
4-12.

2. Using the Auxiliary I/O cable, connect the

682XXB/683XXB rear panel

AUX I/O

connector to

the 562 Network Analyzer AUX I/O connector.

3. Using the GPIB cable, connect the 562 Network

Analyzer DEDICATED GPIB connector to the
682XXB/683XXB

IEEE-488 GPIB

connector.

4. Connect the RF Detector to the 562 Network

Analyzer Channel A Input connector.

5. Connect the 682XXB/683XXB

RF OUTPUT

con-

nector to the RF Detector via a 10 dB Attenuator.

NOTE

Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.

682XXB/683XXB MM

4-17

Serial

I/O

682XXB / 683XXB

COM1

or

COM2

IBM-Compatible PC

562 Network Analyzer

10 dB

Attenuator

GPIB

AUX

I/O

RF Detector

Dedicated

GPIB

RF

Output

A

AUX

I/O

Figure 4-3.

Equipment Setup for Switched Filter Shaper Calibration

SWITCHED FILTER

CALIBRATION

SHAPER CALIBRATION