Alc slope dac adjustment 4-24, Alc slope dac adjustment -24, Alc slope calibration calibration – Anritsu 682XXB User Manual

Page 134

Advertising
background image

NOTE

Before beginning this calibration proce-
dure, always let the 683XXB warm up for a
minimum of one hour.

ALC Slope
DAC
Adjustment

The following procedure lets you adjust the ALC
Slope DAC to compensate for decreasing output
power-vs-frequency for frequencies

£2 GHz and for

frequencies >2 GHz.

1. Set up the 562 Network Analyzer as follows:

a. Press the System Menu key.

b. From the System Menu display, select RESET.

c. Press CHANNEL 2 DISPLAY: OFF.

d. Press CHANNEL 1 DISPLAY: ON.

e. Press CHANNEL 1 MENU key.

f. From the Channel 1 Menu display, select

TRANSMISSION and SELECT INPUT
(NON-RATIO A).

Note

For instruments that do not produce fre-
quencies of <2 GHz, skip to step 6.

2. Set up the 683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

then

Reset

. Upon reset the CW Menu is dis-

played.

b. Press

Edit L1

and set L1 to 5 dB less than

maximum leveled power for the instrument be-
ing calibrated (refer to Table 3-2, page 3-6).

c. Press

Step

. The Step Sweep Menu is dis-

played.

d. Press

Edit F1

and set the F1 start frequency

to 0.01 GHz.

e. Press

Edit F2

and set the F2 stop frequency to

2.1 GHz.

f. Press

Sweep Ramp

. At the resulting Step

Sweep Ramp Menu, press

Num of Steps

and

set the number of steps to 200.

g. Press

Step

to return to the Step Sweep Menu

display.

4-24

682XXB/683XXB MM

ALC SLOPE

CALIBRATION

CALIBRATION