This procedure provides the steps necessary to perform ALC Band-
width calibration. The ALC Bandwidth is adjusted to compensate for
gain variations of the modulator. The adjustment is performed for
each frequency band. This provides a more consistent bandwith
throughout the frequency range of the instrument.

Equipment
Setup

Connect the equipment, shown in Figure 4-9, as fol-
lows:

1. Interface the PC to the 682XXB/683XXB by per-

forming the initial setup procedure, pages 4-7 to
4-12.

NOTE

Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.

Bandwidth
Calibration

The following procedure lets you (1) calibrate the
ALC bandwith and (2) store the calibration data in
non-volatile memory (EEPROMs) on the A17 CPU
PCB.

1. Enter the ALC Bandwidth calibration routine as

follows:

a. At the

$

prompt on the PC display, type:

calterm 110

and press <ENTER>.

The following will appear on the PC display:

Enter calculated gain X.XX

Input:

b. Type:

1.50

and press <ENTER>.

682XXB/683XXB MM

4-31

ALC BANDWIDTH

CALIBRATION

CALIBRATION

Serial

I/O

682XXB / 683XXB

COM1

or

COM2

IBM-Compatible PC

Figure 4-9.

Equipment Setup for AM Bandwidth Calibration