AM CALIBRATION

This procedure provides the steps necessary to perform AM calibra-
tion. This consists of calibrating the AM Calibration DAC, the AM Me-
ter circuit, and the AM Function Generator. The AM Calibration DAC
is calibrated for input sensitivities of 100%/V (linear mode) and
25 dB/V (logarithmic mode) for frequencies

£2 GHz and >2 GHz.

Equipment
Setup

Connect the equipment, shown in Figure 4-10, as
follows:

1. Interface the PC to the 682XXB/683XXB by per-

forming the initial setup procedure, pages 4-7 to
4-12.

2. Connect the Function Generator Output to the

BNC tee. Connect one leg of the tee to the
682XXB/683XXB front panel

AM IN

. Connect the

other leg of the tee to the DMM input.

3. Calibrate the Power Meter with the Power Sen-

sor.

682XXB/683XXB MM

4-33

AM

CALIBRATION

CALIBRATION

RF

OUT

Power Meter

Power

Sensor

Serial

I/O

682XXB / 683XXB

COM1

or

COM2

IBM-Compatible PC

Function

Generator

OUTPUT

AM

IN

BNC
Tee

INPUT

DMM

Figure 4-10.

Equipment Setup for AM Calibration