Fm calibration procedure 4-38, Fm calibration procedure -38, Fm calibration calibration – Anritsu 682XXB User Manual

Page 148

Advertising
background image

4. Connect the 682XXB/683XXB

RF OUTPUT

to the

Spectrum Analyzer RF Input.

NOTE

Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.

FM
Calibration
Procedure

The following steps in the procedure lets you (1)
calibrate the FM Meter circuit, (2) calibrate the FM
Offset Balance DAC, (3) adjust the FM Gain Control
DAC in both narrow and wide modes, and (4) store
the results in non-volatile memory (EEPROM) on
the A17 CPU PCB.

NOTE

To ensure accurate calibration, each step of
this procedure must be performed in se-
quence.

1. Perform FM Meter calibration as follows:

a. At the

$

prompt on the PC screen, type:

calterm 123

and press <ENTER>.

b. Set up the Function Generator for a 83.3 kHz

sinewave with an output level of 0.707 volts
RMS (2 volts peak to peak). When done, press
any key to continue calibration.

The

$

prompt will appear on the screen when

the calibration is complete.

c. Record step completion on the Test Record.

2. Perform FM Offset Balance calibration as follows:

a. At the

$

prompt on the PC screen, type:

calterm 126

and press <ENTER>. (The

$

prompt will appear on the screen when the
calibration is complete.)

b. Record step completion on the Test Record.

3. External FM Wide mode sensitivity calibration is

accomplished by adjusting the FM Gain Control
DAC to obtain 200 MHz and 20 MHz FM devia-
tions at frequencies of 5 GHz and 15 GHz. Modu-
lating signal inputs are from the external
Function Generator.

4-38

682XXB/683XXB MM

FM

CALIBRATION

CALIBRATION

NOTE

To save the calibration data after
completing any calibration step,
type:

calterm 787

and press

<ENTER>.