Phase modulation calibration procedure 4-52, Fm calibration procedure -52, Phase modulation calibration ( f m) calibration – Anritsu 682XXB User Manual

Page 162

Advertising
background image

4. Connect the 682XXB/683XXB

RF OUTPUT

to the

Spectrum Analyzer RF Input.

NOTE

Before beginning this calibration proce-
dure, always let the 682XXB/683XXB
warm up for a minimum of one hour.

FM
Calibration
Procedure

The following steps in the procedure lets you cali-
brate the FM Variable Gain Control DAC and the
FM Flatness DAC in both Wide and Narrow FM
modes and store the results in non-volatile memory
(EEPROM) on the A17 CPU PCB.

NOTE

To ensure accurate calibration, each step of
this procedure must be performed in se-
quence.

1. External Wide FM Mode Sensitivity calibration is

accomplished by adjusting the FM Variable Gain
Control DAC to reduce the carrier level as low as
possible at frequencies of 5 GHz and 20 GHz. The
modulating signal input is from the external
Function Generator.

Perform the calibration as follows:

a. At the

$

prompt, type:

calterm 149

and

press <ENTER>.

b. Set up the Function Generator for a 99.8 kHz

sine wave with an output level of 0.707 volts
RMS (2 volts peak to peak). Use a frequency
counter to verify the output frequency of your
function generator is set to

±1%.

c. On the Spectrum Analyzer, set the Span/Div to

50 kHz per division.

d. On the computer keyboard, use the ‘, 1, 2, and

3 keys to increment and the 7, 8, 9, and 0 keys
to decrement the value of the DAC’s setting.
Start calibration by pressing an increment key.

e. While observing the first Bessel null (Figure

4-16) on the Spectrum Analyzer display, adjust
the value of the DAC’s setting to reduce the
carrier level as low as possible.

f. When finished setting the DAC, press

Q

on the

keyboard to go to the next calibration step.

4-52

682XXB/683XXB MM

PHASE MODULATION

CALIBRATION

(

FM) CALIBRATION

CENTER FREQUENCY

BESSEL

NULL

Figure 4-16.

Typical Spectrum Analyzer Display

of Bessel Null on

FM Waveform

NOTE

You may need to adjust the RBW
setting on the Spectrum Analyzer in
order to see the >–40 dBc null.

Advertising
This manual is related to the following products: