12 recommended test equipment 1-12, Recommended test equipment, 1-12, Test equipment, recommended, 1-12 – Anritsu 682XXB User Manual

Page 22: Recommended test equipment -12, General recommended information test equipment

Advertising
background image

1-12

RECOMMENDED TEST
EQUIPMENT

Table 1-2 provides a list of recommended test equipment needed for
the performance verification, calibration, and troubleshooting proce-
dures presented in this manual.

1-12

682XXB/683XXB MM

INSTRUMENT

CRITICAL

SPECIFICATION

RECOMMENDED

MANUFACTURER/MODEL

USAGE

(1)

Spectrum Analyzer,
with
External Mixers
and
Diplexer Assy

Frequency Range: 0.01 to 65 GHz
Resolution Bandwidth: 10 Hz

Tektronix, Model 2794, with
External Mixers:

WM780K (18 to 26.5 GHz)
WM780A (26.5 to 40 GHz)
WM780U (40 to 60 GHz)
WM780E (60 to 90 GHz)

Diplexer Assy: 015-0385-00

C, P, T

Spectrum Analyzer

Frequency Range: 20 Hz to 40 MHz
Resolution Bandwidth: 3 Hz

Hewlett-Packard, Model 3585B

P

Frequency Counter
with
Cable Kit
and
External Mixer

Frequency Range: 0.01 to 65 GHz
Input Impedance: 50

W

Resolution: 1 Hz
Other: External Time Base Input

EIP Microwave, Inc. Models 538B, 548B,
or 578B, with
Cable Kit: Option 590
and
External Mixer:

Option 91 (26.5 to 40 GHz)
Option 92 (40 to 60 GHz)
Option 93 (60 to 90 GHz)

C, P

Power Meter,
with Power Sensor

Power Range: –30 to +20 dBm

(1

m

W to 100mW)

Hewlett-Packard Model 437B, with
Power Sensor:

HP 8487A (0.01 to 50 GHz)

C, P

Power Meter,
with Power Sensor

Power Range: –30 to +20 dBm

(1

m

W to 100mW)

ANRITSU ML4803A, with
Power Sensor:

MP716A4 (50 to 75 GHz)

C, P

Digital Multimeter

Resolution: 4-1/2 digits (to 20V)
DC Accuracy: 0.002% +2 counts
DC Input Impedance: 10 M

W

AC Accuracy: 0.07% +100 counts

(to 20 kHz)

AC Input Impedance: 1 M

W

John Fluke, Inc., Model 8840A, with
Option 8840A-09K (True RMS AC)

C, T

Frequency Reference

Frequency: 10 MHz
Accuracy: 5 x 10

-12

parts/day

Absolute Time Corp., Model 300

P

Frequency Standard

Frequency: 10 MHz
Accuracy: 1

´

10

–10

parts/day

Spectracom Corp., Model 8161

C

Function Generator

Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms

Hewlett-Packard, Model 8116A

C

Function Generator

Output Voltage: 2 volts peak-to-peak
Functions: 0.1 Hz to 100 kHz sine and
square waveforms

Hewlett-Packard, Model 33120A

C

Table 1-2.

Recommended Test Equipment (1 of 3)

GENERAL

RECOMMENDED

INFORMATION

TEST EQUIPMENT

Advertising