6-12

REMOVING AND
REPLACING THE
A21-1 PCB

This paragraph provides instructions for removing and replacing the
A21-1 BNC/AUX I/O Connector PCB, which is located on the rear
panel assembly (see Figure 6-4).

Preliminary

Disconnect the power cord from the unit. Remove
the chassis covers as described in paragraph 6-2.
Remove the rear panel assembly as described in
paragraph 6-10.

WARNING

When power is removed from the instrument, always allow
five minutes for the capacitors on the A21 Line Filter/Recti-
fier PCB to discharge before removing the A21-1 PCB.

Procedure

Remove and replace the A21-1 PCB as follows:

Step 1

Using a Phillips screwdriver, remove the
four screws that fasten the A21 shield to
the A21 PCB.

Step 2

Remove the shield and set it aside.

Step 3

Disconnect the ribbon cable connector
from the A21-1 PCB.

Step 4

Using a

ANRITSU

P/N T1451 tool, remove

the dress nuts from the 12 rear panel
BNC connectors.

Step 5

Carefully remove the A21-1 PCB from the
rear panel assembly.

Step 6

To replace the PCB, reverse the removal
process.

682XXB/683XXB MM

6-17

REMOVAL AND
REPLACEMENT PROCEDURES

A21-1 PCB

Advertising