13 removing and replacing the fan assembly 6-18, Preliminary 6-18, Procedure 6-18 – Anritsu 682XXB User Manual

Page 238: Remove/replace procedures, 6-18, Fan assembly, 6-18, Removing and replacing the fan, Assembly -18, Preliminary -18 procedure -18, Removal and fan replacement procedures assembly

Advertising
background image

6-13

REMOVING AND
REPLACING THE FAN
ASSEMBLY

This paragraph provides instructions for removing and replacing the
fan assembly, which is located on the rear panel assembly (see Figure
6-4).

Preliminary

Disconnect the power cord from the unit. Remove
the chassis covers as described in paragraph 6-2.
Remove the rear panel assembly as described in
paragraph 6-10.

WARNING

When power is removed from the instrument, always allow
five minutes for the capacitors on the A21 Line Filter/Recti-
fier PCB to discharge before removing the rear panel assem-
bly.

Procedure

Remove and replace the fan assembly as follows:

Step 1

With the rear panel laying flat, use a
Phillips screwdriver to remove the four
screws and flat washers that fasten the
fan mount to the rear panel.

Step 2

Lift the fan mount, containing the fan as-
sembly, from the rear panel assembly.

Step 3

Remove the fan assembly from the fan
mount.

Step 4

Clean the honeycomb fan filter as re-
quired before replacing the fan assembly.

Step 5

To replace the fan assembly, reverse the
removal process.

NOTE

To ensure proper cooling of the unit,
always mount the fan assembly
with the airflow direction indicator
arrow on the fan body pointing to-
ward the interior of the instrument.

6-18

682XXB/683XXB MM

REMOVAL AND

FAN

REPLACEMENT PROCEDURES

ASSEMBLY