Test procedure 3-9, Test procedure -9 – Anritsu 682XXB User Manual

Page 69

Advertising
background image

Test
Procedure

The frequency error is measured at the start and
finish of the test time period of 24 hours. The aging
rate is the difference between the two error read-
ings.

1. Set up the Frequency Reference as follows:

a. Press the

ESC

key until the MAIN MENU is

displayed.

b. At the MAIN MENU display, press

1

to select

CONFIGURATION.

c. At the CONFIGURATION MENU display,

press

8

to select MEAS.

d. Press the

MOD

key and use the Up/Down Ar-

row keys to get to the menu display:
MEASUREMENT = FREQ.

e. Press the

ENTER

key.

f. Press the

ESC

key until the MAIN MENU is

displayed.

g. At the MAIN MENU display, press

3

to select

the REVIEW MENU.

h. At the REVIEW MENU display, press

8

to se-

lect TFM.

2. Connect the 682XXB/683XXB rear panel

10 MHz

REF OUT

signal to the Frequency Reference front

panel 10 MHz input.

3. Wait approximately 90 minutes (default setting)

until the FMFOM on the Frequency Reference
display decreases from 9 to 1. (The default setting
is recommended to achieve optimum mease-
ments.)

The frequency error in the signal under test is
displayed in ps/s (Picoseconds/Second). For exam-
ple, an error of –644681 ps/s is –644681

´ 10

–12

or

–6.44681

´ 10

–7

away from the 10 MHz internal

reference of the Frequency Reference.

The frequency error display is continuously up-
dated as a running 5000-second average. The av-
eraging smooths out the short-term instability of
the oscillator.

682XXB/683XXB MM

3-9

PERFORMANCE

INTERNAL TIME BASE

VERIFICATION

AGING RATE TEST