7 frequency synthesis tests 3-11, Test setup 3-11, Frequency synthesis tests, 3-11 – Anritsu 682XXB User Manual

Page 71: Frequency synthesis tests -11, Test setup -11, Performance frequency verification synthesis tests

Advertising
background image

3-7

FREQUENCY

SYNTHESIS TESTS

The following tests can be used to verify correct operation of the fre-
quency synthesis circuits. Frequency synthesis testing is divided into
two parts—coarse loop/YIG loop tests and fine loop tests.

Test Setup

Connect the equipment, shown in Figure 3-2, as fol-
lows:

1. Connect the 682XXB/683XXB rear panel

10 MHz

REF OUT

to the Frequency Counter 10 MHz Ex-

ternal Reference input. If the Frequency Counter
has an INT/EXT toggle switch, ensure the switch
is set to EXT.

2. Connect the 682XXB/683XXB

RF OUTPUT

to the

Frequency Counter RF Input as follows:

a. For measuring frequencies of 0.01 to 1.0 GHz,

connect to the Band 2 input (Connection A).

b. For measuring frequencies of 1.0 to 26.5 GHz,

connect to the Band 3 input (Connection A).

c. For measuring frequencies of 26.5 to 40.0 GHz,

connect to the Band 4 input via the Option 91
waveguide mixer (Connection B).

d. For measuring frequencies of 40.0 to 60.0 GHz,

connect to the Band 4 input via the Option 92
waveguide mixer (Connection B).

e. For measuring frequencies of 60.0 to 65.0 GHz,

connect to the Band 4 input via the Option 93
waveguide mixer (Connection B).

682XXB/683XXB MM

3-11

PERFORMANCE

FREQUENCY

VERIFICATION

SYNTHESIS TESTS

682XXB/683XXB

FREQUENCY COUNTER

RF IN

Band 2

Input

Band 3

Input

Connection A

(0.01 to 26.5 GHz)

Connection B

(26.5 to 60 GHz)

10 MHz

EXT IN

Band 4

Input

Mixer

10 MHz

REF OUT

RF OUT

Figure 3-2.

Equipment Setup for Frequency Synthesis Tests