Fine loop test procedure 3-13, Fine loop test procedure -13, Performance frequency verification synthesis tests – Anritsu 682XXB User Manual

Page 73

Advertising
background image

Fine Loop
Test
Procedure

The following procedure tests the fine loop by step-
ping the instrument through ten 1 kHz steps (ten
100 Hz steps for instruments with Option 11) and
measuring the RF output at each step.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to the first test frequency indicated on

the Test Record.

2. Record the Frequency Counter reading on the

Test Record. The Frequency Counter reading
must be within

±100 Hz of the displayed 682XXB/

683XXB frequency (

±10 Hz for instruments with

Option 11) to accurately complete this test.

3. On the 682XXB/683XXB, use the cursor control

key (diamond-shaped key) to increment F1 to the
next test frequency on the Test Record. Record
the Frequency Counter reading on the Test Re-
cord.

4. Repeat step 3 until all frequencies listed on the

Test Record have been recorded.

682XXB/683XXB MM

3-13

PERFORMANCE

FREQUENCY

VERIFICATION

SYNTHESIS TESTS