Test setup 3-18, Harmonic test: rf output 2 to 20 ghz, 3-18, Harmonic test: rf output signals – Anritsu 682XXB User Manual

Page 78: From 2 to 20 ghz -18, Test setup -18, Spectrum analyzer

Advertising
background image

3-9

HARMONIC TEST: RF
OUTPUT SIGNALS FROM
2 TO 20 GHz

The following test can be used to verify that the 682XXB/683XXB
meets its harmonic specifications for RF output signals from 2 to
20 GHz (2.2 to 20 GHz for 68XX5B Models). Test record entries are
supplied for harmonics up to a frequency limit of 40 GHz. Additional
harmonic checks may be made at any frequency of interest up to the
RF output frequency limit of the 682XXB/683XXB model being tested.
These additional harmonic checks can be accomplished through the
use of waveguide mixers to extend the frequency range of the spec-
trum analyzer.

Test Setup

Connect the equipment, shown in Figure 3-4, as fol-
lows:

1. Connect the 682XXB/683XXB rear panel

10 MHz

REF OUT

to the Spectrum Analyzer External

Reference Input.

2. Connect the diplexer and appropriate external

waveguide mixer to the Spectrum Analyzer.

3. Connect the 682XXB/683XXB

RF OUTPUT

to the

Spectrum Analyzer as shown in Connection A
(682XXB/683XXB

RF OUTPUT

to Spectrum

Analyzer RF IN).

3-18

682XXB/683XXB MM

PERFORMANCE

HARMONIC TEST: RF OUTPUT

VERIFICATION

SIGNALS FROM 2 TO 20 GHz

10 MHz

REF OUT

682XXB / 683XXB

EXT REF

INPUT

Mixer

Diplexer

Connection B

RF OUT

Spectrum Analyzer

RF IN

Connection A

Figure 3-4.

Equipment Setup for Harmonic Test: RF Output Signals from 2 to 20 GHz

Advertising