Power level accuracy test procedure 3-27, Power level flatness test procedure 3-27 – Anritsu 682XXB User Manual

Page 87

Advertising
background image

Power Level
Accuracy Test
Procedure

Power level accuracy is tested by stepping the out-
put power level down in 1 dB increments from its
maximum rated power level and measuring the out-
put power level at each step.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to the CW frequency indicated on the

Test Record.

d. Press

Edit L1

to open the current power level

parameter for editing.

e. Set L1 to the power level indicated on the Test

Record.

2. Measure the output power level with the Power

Meter and record the reading on the Test Record.

3. On the 682XXB/683XXB, use the cursor control

key (diamond-shaped key) to decrement L1 to the
next test power level on the Test Record. Measure
and record the Power Meter reading on the Test
Record.

4. Repeat step 3 for each of the test power levels

listed on the Test Record for the current CW fre-
quency.

5. Repeat steps 1 thru 4 for all CW frequencies

listed on the Test Record.

Power Level
Flatness Test
Procedure

Power level flatness is tested by measuring the out-
put power level variation during a full band sweep;
first in the step sweep mode (682XXB/683XXB mod-
els), then in analog sweep mode (683XXB models
only).

1. Set up the 682XXB/683XXB as follows for a step

sweep power level flatness test:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. The CW Menu is displayed.

b. Press

Step

to place the signal generator in the

step sweep frequency mode and display the
Step Sweep Menu.

682XXB/683XXB MM

3-27

PERFORMANCE

POWER LEVEL ACCURACY

VERIFICATION

AND FLATNESS TESTS