AMPLITUDE
MODULATION TEST

This procedure verifies the operation of the 682XXB/683XXB ampli-
tude modulation input sensitivity circuit.

The RF OUTPUT of the 682XXB/683XXB is monitored on a Spectrum
Analyzer display. The (modulated) IF Output of the Spectrum Ana-
lyzer is monitored with a Modulation Analyzer. A 50% AM (default
value) signal is set to a reference point on the Spectrum Analyzer. The
actual modulation value is then computed from the indicated Modula-
tion Meter values. (The absolute values of the AM PK+ and AM PK–
readings are used in the given formula to compensate for non-linearity
errors in the test equipment.)

Test Setup

Connect the equipment, shown in Figure 3-7, as fol-
lows:

1. Connect the 682XXB/683XXB rear panel 10 MHz

REF OUT to the Spectrum Analyzer External
Reference Input.

2. Connect the IF Output of the Spectrum Analyzer

to the RF Input of the Modulation Analyzer.

3. Connect the RF OUTPUT of the 682XXB/683XXB

to the Spectrum Analyzer RF Input.

3-30

682XXB/683XXB MM

682XXB / 683XXB

RF

OUTPUT

10 MHz
REF OUT

EXT

REF Input

RF Input

RF Input

IF Output

Modulation

Analyzer

Spectrum

Analyzer

Figure 3-7.

Equipment Setup for Amplitude Modulation Tests

PERFORMANCE

AMPLITUDE

VERIFICATION

MODULATION TEST