AM Input
Sensitivity
Procedure

The following procedure lets you measure the abso-
lute peak AM values for a 50% AM signal and calcu-
late the actual modulation value.

1. Set up the 682XXB/683XXB as follows:

a. Reset the instrument by pressing

SYSTEM

,

then

Reset

. Upon reset, the CW Menu is dis-

played.

b. Press

Edit F1

to open the current frequency

parameter for editing.

c. Set F1 to 5.0 GHz.

d. Press

Edit L1

to open the current power level

parameter for editing.

e. Set L1 to a power level that is 6 dB below

maximum rated power (refer to Table 3-2, page
3-6).

f. Press

MODULATION

, then

AM

. At the result-

ing AM Status menu, press

On/Off

to turn AM

on.

2. Set up the Spectrum Analyzer as follows:

a. CF: 5.0 GHz

b. Span/Div: 0 Hz

c. RBW: 100 kHz

d. MIN Noise: Activate

3. On the Spectrum Analyzer, adjust the reference

level to place the trace 6 to 8 dB below the top
graticule of the display.

4. Set up the Modulation Analyzer for:

a. AM PK(+)

b. 300 Hz High-Pass Filter

c. 20 kHz Low-Pass Filter

5. Measure the peak AM on the Modulation Ana-

lyzer. Record the AM PK(+) reading on the Test
Record.

6. Press PK(–) on the Modulation Analyzer.

7. Measure the peak AM on the Modulation Ana-

lyzer. Record the AM PK(–) reading on the Test
Record.

682XXB/683XXB MM

3-31

PERFORMANCE

AMPLITUDE

VERIFICATION

MODULATION TEST