This procedure verifies the operation of the 682XXB/683XXB fre-
quency modulation input sensitivity circuitry.

The RF OUTPUT of the 682XXB/683XXB is monitored on a Spectrum
Analyzer display. In Unlocked Wide FM mode, the Spectrum Analyzer
is used as a frequency counter to measure the actual FM deviation and
determine FM accuracy. In Unlocked Narrow, Locked, and Locked
Low-Noise FM modes, the Spectrum Analyzer is used to monitor the
carrier level. FM accuracy is determined by measuring the actual FM
deviation necessary to reduce the carrier level to its lowest level.
These tests qualify how the FM input signal affects the signal genera-
tor’s RF output.

Test Setup

Connect the equipment, shown in Figure 3-8, as fol-
lows:

1. Connect the 682XXB/683XXB rear panel

10 MHz

REF OUT

to the Spectrum Analyzer External Ref-

erence Input.

2. Connect the

RF OUTPUT

of the 682XXB/683XXB

to the Spectrum Analyzer RF Input.

FM Input
Sensitivity
Procedure

The following procedure lets you measure the FM
deviation for a specific input sensitivity in Unlocked
Wide, Unlocked Narrow, Locked, and Locked Low-
Noise FM modes and determine FM accuracy.

682XXB/683XXB MM

3-33

10 MHz

REF OUT

682XXB / 683XXB

EXT REF

INPUT

RF OUT

Spectrum Analyzer

RF IN

Figure 3-8.

Equipment Setup for Frequency Modulation Tests

PERFORMANCE

FREQUENCY

VERIFICATION

MODULATION TESTS