3-14

PULSE MODULATION
TESTS: RISE TIME,
FALL TIME, OVER-
SHOOT, AND LEVEL

Pulse modulation tests verify correct operation as
well as rise time, fall time, overshoot, and level. The
pulse leveling accuracy test compares the pulsed RF
output level to verify the performance of the level
detector(s), detector amplifiers, and ALC sample/
hold circuits.

Test Setup

Connect the equipment, shown in Figure 3-10, as
follows:

1. Connect the 682XXB/683XXB rear panel

PULSE

SYNC OUT

to the Oscilloscope Trigger Input.

2. Connect the Pulse Detector Output to the Vertical

Input of the Oscilloscope.

3. On the Oscilloscope, select 50

W input impedance.

4. Connect the Pulse Detector Input through a

150 MHz High Pass Filter to the 682XXB/
683XXB

RF OUTPUT

.

682XXB/683XXB MM

3-39

PERFORMANCE

PULSE MODULATION TESTS:

VERIFICATION

RISE TIME, FALL TIME, OVERSHOOT, AND LEVEL

RF Out

Pulse

Detector

150 MHz HPF
(For Overshoot
Measurement)

CH. 4 Trigger

682XXB / 683XXB

Pulse

Sync Out

CH. 2 Input

Oscilloscope

Figure 3-10.

Equipment Setup for Pulse Modulation Tests: Rise Time, Fall Time, Overshoot, and Level