4 grounding, Grounding -4, Important – Yokogawa YVP110 User Manual

Page 33: Warning

Advertising
background image

<4. Wiring and Piping>

4-4

IM 21B04C01-01E

1) Screw the flameproof packing adapter into the

terminal box until the O-ring touches the wiring

port (at least 6 full turns), and firmly tighten the

lock nut.

2) Insert the cable through the union cover, the

union coupling, the clamp nut, the clamp ring,

the gland, the washer, the rubber packing, and

the packing box, in that order.

3) Insert the end of the cable into the terminal box.

4) Tighten the union cover to grip the cable.

When tightening the union cover, tighten

approximately one turn past the point where the

cable will no longer move up and down.

IMPORTANT

Proper tightening is important. If it is too tight, a

circuit break in the cable may occur; if not tight

enough, the flameproof effectiveness will be

compromised.

5) Fasten the cable by tightening the clamp nut.

6) Tighten the lock nut on the union cover.

7) Connect the cable wires to each terminal.

O-ring

Adapter body

Lock nut

Wrench

Packing box

Rubber packing

Washer

Gland

Clamp ring

Clamp nut
Union coupling
Lock nut
Wrench
Union cover
Cable

Apply a non-hardnening

sealant to the threads for

waterproofing.

F0405.ai

Figure 4.5

Installing Flameproof Packing Adapter

Flameproof metal conduit wiring

• A seal fitting must be installed near the

terminal box connection port for a sealed

construction.

• Apply a non-hardening sealant to the

threads of the terminal box connection port,

flexible metal conduit and seal fitting for

waterproofing.

F0406.ai

Non-hazardous area

Hazardous area

Flameproof

heavy-gauge

steel conduit

Tee

Drain plug

Seal fitting

Gas sealing device

Flameproof flexible

metal conduit

Apply a non-hardening

sealant to the threads of

these fittings for

waterproofing

After wiring, impregnate the fitting

with a compound to seal tubing.

Figure 4.6

Typical Wiring Using Flameproof Metal

Conduit

4.4 Grounding

Grounding is always required for the proper

operation of transmitters. Follow the domestic

electrical requirements as regulated in each

country.
Ground terminals are located on the inside and

outside of the terminal box. Either of these terminals

may be used. See Figure 4.2.

WARNING

For JIS flameproof type and intrinsically

safe type, grounding should satisfy Class D

requirements (grounding resistance, 100 Ω or

less).