2 auto calibration, 2 auto calibration -2, Operating procedures – Yokogawa DAQWORX User Manual
Page 175
Advertising

8-2
IM WX102-01E
8.2
Auto Calibration
Operating Procedures
1
Click the connect button on the calibration screen. The DA100, DC100, or DR
system that is connected will be displayed.
Module that does not need to be calibrated
‘
’ marks are displayed for modules that do not need to be calibrated.
2
Clicking the module to be calibrated displays the calibration dialog box.
Auto tab
Select all
ranges
Range
Select the terminal
type
Start button
3
Click the Auto tab to display the auto calibration screen.
4
Clicking the range to be calibrated shows the selected range in red. To calibrate
all ranges click All.