Yokogawa DAQWORX User Manual

Page 181

Advertising
background image

8-8

IM WX102-01E

Set the Calibration Value to a Previously Decided Value

1.

Click the connection button in the calibration screen. The system of connected

DA100s is displayed. A [

] mark is displayed on modules that don’t require

calibration.

2.

When you click a module to be calibrated, the calibration dialog box appears.

3.

Click a terminal to be calibrated (CH1, CH2, etc.) the selected terminal is

displayed in red. When calibrating the entire range, click [All].

4.

Enter the calibration value for each Zero and FULL. When you click the [Zero] and

[FULL] buttons, the setting range is displayed.

5.

When you click the [start] button, a message appears to confirm that you wish to

save the calibration values.

6.

When you click [OK], a message appears to reconfirm that you wish to save the

current calibration values.

7.

When you click [OK], a report appears telling you that you calibration values were

accepted.

8.

Click [OK] to conclude calibration. When calibrating a separate module, click

the [Close] button, and repeat the procedure from step 2. [OK] is displayed on

modules for which calibrating has been completed.

8.4 Calibrating the Transmission Output Module (DA, DC only)