Computation (uniquely for da100 with /m1 option) – Yokogawa DAQWORX User Manual

Page 26

Advertising
background image

1-18

IM WX102-01E

• Math computation start/stop/clear/reset

Clear:

When issued during the computation, the measured data is reset before doing the
first computation.

Computation clear

Computed data

Computation start

Measurement interval

Reset:

When issued during the computation, the measured data is reset after doing the
first computation.

Measurement interval

Computation start

Computation

reset

Computed data

• Report start/stop
For models with Report option it is possible to start/stop reporting.

Computation (Uniquely for DA100 with /M1 option)

The DA100 can execute computations with the measured data of each input channel

taken as a variable. The results can be displayed or saved. The following operators can

be used for computation.

Basic operators

Type

Operator

Example

Description

Addition

+

001+002

To obtain the sum of the measured data of channel 001
and channel 002.

Subtraction

-

002-001

To obtain the difference of the measured data of
channel 002 and channel 001.

Multiplication

*

003*K1

To multiply constant K1 to the measured data of channel
003.

Division

/

004/K2

To divide the measured data of channel 004 by constant
K2.

Power

**

005**006

To take the power of measured data of channel 005 with
the measured data of channel 006.

Absolute
value

ABS()

ABS(001)

To obtain the absolute value of the measured data of
channel 001.

Square root

SQR()

SQR(002)

To obtain the square root of the measured data of
channel 002.

Common
logarithm

LOG()

LOG(003)

To obtain the common logarithm of the measured data
of channel 003.

Natural
Logarithm

LN()

LN(004)

To obtain the natural logarithm of the measured data of
channel 004.

Exponent

EXP()

EXP(005)

To make the measured data of channel 005 to be x and
obtain e

x

.

* +/- can be used as signs as in -(001).

1.5 DA100 Functions