Before Operation

• Resistance temperature detector (RTD)

RTD Type

Display

Measurable Range

Pt100:1mA

PT1

–200.0 to 600.0°C

Pt100:2mA

PT2

–200.0 to 250.0°C

JPt100:1mAJ

JPT1

–200.0 to 550.0°C

JPt100: 2mAJ

JPT2

–200.0 to 250.0°C

Pt50: 2mAJ

PT50

–200.0 to 550.0°C

Ni100:1mA-S

*1

NI1

–200.0 to 250.0°C

Ni100:1mA-D

*2

NI2

–60.0 to 180.0°C

Ni120:1mA

*3

NI3

–70.0 to 200.0°C

Cu10: GE

*4

CU1

–200.0 to 300.0°C

Cu10: L&N

*4

CU2

–200.0 to 300.0°C

Cu10: WEED

*4

CU3

–200.0 to 300.0°C

Cu10: BAILEY

*4

CU4

–200.0 to 300.0°C

Pt100:1mA-H

PT1S

–140.00 to 150.00°C

Pt100: 2mA-H

PT2S

–70.00 to 70.00°C

JPt100:1mA-H

JPT1S

–140.00 to 150.00°C

JPt100: 2mA-H

JPT2S

–70.00 to 70.00°C

J263*B

J263B

–0.0 to 300.0K

*1 RTD (SAMA)

*2 RTD (DIN)

*3 RTD (McGROW EDISON COMPANY)

*4 RTD (Cuid)

Guaranteed accuracy range

Cu10: GE

–84.4 to 170.0°C

Cu10: L&N

–75.0 to 150.0°C

Cu10: WEED

–20.0 to 250.0°C

Cu10: BAILEY

–20.0 to 250.0°C

• Contact point (DI)

Input Type

Display

Measurable Range

VOLT

LEVL

0 to 1

*1

CONTACT

CONT

0 to 1

*2

*1 OFF if below 2.4V, ON if above 2.4V
*2 Contact point ON/OFF

• mA input (mA)

Measurement Range

Display

Measurable Range

20mA

20mA

–20.000 to 20.000mA

• Pulse input (PULSE)

Measurement Range

Measurable Range

RATE

Determine the count number during 1s interval.

GATE

Determine the most occurring state, ON or OFF, during 1s
interval.

• Power monitor (AC)

Input Range

Display

250V-0.5A

250V-0.5A

250V-5A

250V-5A

25V-0.5A

25V-0.5A

25V-5A

250V-5A

• Strain input (STRAIN)

Measurement Range

Display

Measurable Range

2k

2k

–2000.0 to 2000.0µε

20k

20k

–20000 to 20000µε

200k

200k

–20000 to 20000µε