Setting the span (span), Precautions in power module measurement – Yokogawa DAQWORX User Manual

Page 57

Advertising
background image

4-10

IM WX102-01E

Setting the Span (SPAN)

Set the left and right spans within the limits of an input range. In the SPAN menu item,

the left-hand value is the left span and the right-hand value the right span. Set the span

within the measuring range. The measurable limits vary depending on the measuring

range you select, as shown below.

25V-0.5A

25V-5A

250V-0.5A

250V-5A

Effective voltage
Vi (i=1,2,3,13,0)

0.00 to 25.00Vrms

0.00 to 25.00Vrms

0.0 to 250.0Vrms

0.0 to 250.0Vrms

Effective current
Ii (i=1,2,3,13,0)

0.0000 to 0.5000Arms

0.000 to 5.000Arms

0.0000 to 0.5000Arms

0.000 to 5.000Arms

Active power
P1,P2,P3

–12.50 to 12.50W

–125.0 to 125.0W

–125.0 to 125.0W

–1250 to 1250W

Active power
P13

–25.00 to 25.00W

–250.0 to 250.0W

–250.0 to 250.0W

–2500 to 2500W

Active power
P0

–37.50 to 37.50W

–375.0 to 375.0W

–375.0 to 375.0W

–3750 to 3750W

Apparent power
VA1,VA2,VA3

0.00 to 12.50VA

0.0 to 125.0VA

0.0 to 125.0VA

0 to 1250VA

Apparent power
VA13

0.00 to 25.00VA

0.0 to 250.0VA

0.0 to 250.0VA

0 to 2500VA

Apparent power
VA0

0.00 to 37.50VA

0.0 to 375.0VA

0.0 to 375.0VA

0 to 3750VA

Reactive power
Var1,Var2,Var3

0.00 to 12.50Var

0.0 to 125.0Var

0.0 to 125.0Var

0 to 1250Var

Reactive power
Var13

0.00 to 25.00Var

0.0 to 250.0Var

0.0 to 250.0Var

0 to 2500Var

Reactive power
Var0

0.00 to 37.50Var

0.0 to 375.0Var

0.0 to 375.0Var

0 to 3750Var

Power factor
PFi(i=1,2,3,13,0)

–1.00 to 1.00

–1.00 to 1.00

–1.00 to 1.00

–1.00 to 1.00

Phase
PHi(i=1,2,3,13,0)

–80.0 to 80.0deg

–80.0 to 80.0deg

–80.0 to 80.0deg

–80.0 to 80.0deg

Frequency
FREQ

45.00 to 65.00Hz

45.00 to 65.00Hz

45.00 to 65.00Hz

45.00 to 65.00Hz

Precautions in Power Module Measurement

Input the voltage or current being measured at a level between 10% and 100% of the

measuring range. The instrument bases its calculations of all other parameters on the

frequency of V1. Extra care must therefore be taken when setting the input level of

V1. If the input level fails to fall within the given limits, there is no guarantee that the

measurement of any other parameters will be reliable.