Index, Error messages <darwin hardware configuration, Messages – Yokogawa DAQWORX User Manual

Page 207

Advertising
background image

11-5

IM WX102-01E

1

2

3

4

5

6

7

8

9

10

11

Index

T

roubleshooting

Error Messages <DARWIN Hardware Configuration>

No.

Message/Corrective action

E3631

Communication not possible during logging.
Terminate Logger’s communication before executing.

E3632

Communication not possible while monitoring data.
Terminate Logger’s communication before executing.

E3633

Connected to unknown model.
Modify the system or change the settings.

E3636

Failed to send command.
Some items could not be set. Check the setup data.

E3637

Can’t reconstruct.
Displayed when the DR stand-alone model is connected.

E3638

Initial balancing failed.
Check the input, initialize, and perform initial balancing again. If it fails again, servicing is needed.

E3639

Incorrect file name!
Check the file name and extension.

E3640

File operation failed.
Check the free space on the PC’s hard drive. Check that the hard disk is operating properly.

E3641

These data were generated by another model.
Select a different file.

E3643

Failed to load data. Some data initialized.
Check the setup data.

E3644

Invalid channels are included in the copy source.
Check the copy source and copy destination settings.

E3645

Please enter an equation.
Enter an equation.

E3646

Unknown symbol is entered. Please change equation.
Check the setup data.

E3647

Invalid channel number is entered. Please change equation.
Check the channel number.

E3648

Incorrect use of parenthesis. Please change equation.
Check the setup data.

E3649

Grammatical error. Please change equation.
Check the setup data.

E3650

Character string is incorrect.
Check the setup data.

E3651

Invalid channel number.
Change the channel number.

E3652

For input range, please keep CH numbers: Left < Right.
Check the setup data.

E3653

Channel number is already being used!
Check the setup data.

Messages

No.

Message/Corrective action

M3404

Invalid Project name. Please enter a new name.

M3405

This Project name is already being used. Please enter a new Project name.

M3406

There must be at least one unlocked Project.

M3407

Can't contain any of the following characters / : , ; * ? " < > |

M3031

Do you want to stop recording ?

M3032

Please stop Measuring before you exit.

M3033

You must Unlock and enter a Password to Exit.

M3231

Send Tags?

M3232

Send Tag IDs?

M3601

Start Receiving?

M3602

Start Sending?

M3603

Reconstruct?

M3604

File is already open. Reopen will cause your changes to be discarded. Do you want to reopen?

M3605

All settings will be initialized!

M3606

SET (Regular) settings will be initialized!

M3607

Add Unit?

M3608

Delete Unit?