Yokogawa DAQWORX User Manual

Page 78

Advertising
background image

4-31

IM WX102-01E

1

2

3

4

5

6

7

8

9

10

11

Index

DAR

WIN Hardware Configurator

MATH (for instruments with /M1 option or FDD option only)

• START: Starts computation.

• STOP: Stops computation.
• RESET: Resets computation channel data at the end of completion of the first

computation in case an event takes place.

• CLEAR: Resets computation channel data immediately in case an event takes

place.

• If MATH is selected as a level action, computation will be carried out while an event

is present. Computation will stop when the event is cleared.

MEMORY (for DC100, and DR with FDD option only)
DR with FDD option

• DATA_WR : Saves a data item each time an event occurs, until the specified data

length is reached.

• WR_TRIG : Saves measured/computed data on the built-in RAM disk.
• RD_TRIG : Reads measured/computed data from the built-in RAM disk.
• LD_TRG1-3 : Reads setup data from the built-in RAM disk.

DC100

• DATA_WR : Writes the data for one scan (one data/channel) retained at the

generation of an event. To execute this, it is necessary to set LOGIC

to the writing period.

• TRIG :

Generates a trigger to start writing. To execute this, it is necessary to

set TRIG to WRITE TRIG.

• START :

Opens a file and starts the writing operation. It is the same function
as the START key on the operation panel.

• PAUSE :

In the case of edge action, writing is suspended at the generation of

an event. In the case of level action, by setting the alarm, remote,

relay or internal switch to the event, writing is suspended at the

detection of an event and resumes at the resetting of the event (see
the following figure). Setting the timer, match time or manual function
key to the event repeats suspension and resumption of data writing

at the generation of each event.

• CP_FDD :

All files in the built-in RAM disk are copied on a floppydisk. Whether
to cancel files after copying or not depends on the setting in the MEM
SET of the SET UP menu. Copying after a conversion to ASCII

format cannot be done.

• CP_SCS0 to CP_SCS7 :

All files in the built-in RAM disk are copied on a MO, ZIP or PD.
Whether to cancel files after copying or not depends on the setting
in the MEM SET of the SET UP menu. Copying after a conversion to

ASCII format cannot be done.

Displayed only on models with the C/5 option. SCS0 to SCS7
represent the SCSI ID numbers. Check the ID numbers of SCSI

devices that are connected, because ID numbers are displayed even

if there are no devices connected to them.

FLAG (only for Math option models; DC style number. 7+, DA100/DR style
number 8 or higher)

• F01 to 16: Set the flag number (F01 to F16) to 1. F01 to F16 are normally 0. Flags

can be placed in computing equations to hold the computed result using a certain

event as a trigger, or reset to 0.

4.14 Event/Action, Match Time, and Group Settings

Advertising
This manual is related to the following products: