Input Related Functions

Remote RJC*(RRJC) ← Can be selected only with optional math
function (/M1 option)

In measuring temperature with thermocouples within the same unit, when the item

to be measured is located at a great distance, you can setup relay terminals near

the item, measure the temperature of the relay terminal section using thermocouples

(reference channel), and use the resultant value as the reference junction

compensation value for the temperature measurement. By connecting a copper wire

between the relay terminal and input terminal of the input module, and a thermocouple

between the DUT and relay terminal, you can measure the temperature of the DUT

without the need for a large amount of expensive thermocouples.

Input module

Thermocouple*

Copper wire

Copper wire

Relay terminal (To be provided by the customer)

Reference channel

* Please use the same type of thermocouple.

Input

terminal

Thermocouple*

Thermocouple*

Measurement period

The measurement period applies to all channels. It can be selected from 0.5, 1, 2, 3, 4, 5,

6, 10, 12, 15, 20, 30 or 60 seconds. However, the settable minimum measurement period
varies as shown below according to the filter ON/OFF setting, type of input module and
A/D integral time.

Integral time : 50/60Hz

Integral time : 100ms

Input module

Filter OFF

Filter ON

Filter OFF

Filter ON

10CH model

0.5s

3s

4s

12s

20CH model

2s

4s

5s

15s

30CH model

2s

4s

6s

20s

A/D integral time

Set the A/D integral time to 100 ms to maximize the noise elimination effect. To shorten
calculation time, select 50Hz, 60Hz or AUTO (50/60Hz switching automatically). The
settable minimum measurement period varies, as shown below according to the A/D
integral time, filter settings and the number of input module channels.
• Integral time: 50/60Hz

Input module

Filter OFF

Filter ON

10CH model

0.5s

3s

20CH model

2s

4s

30CH model

2s

4s

• Integral time: 100ms

Input module

Filter OFF

Filter ON

10CH model

4s

12s

20CH model

5s

15s

30CH model

6s

20s