Computing equations are set according to the following rules:

The number of computing equations

Thirty computing equations for the stand-alone type and sixty for the expandable type

can be set. Each computing equation is assigned a number. The numbers are A01 to

A30 for the stand-alone type and A01 to A60 for the expandable type. These numbers

are called computation channel numbers.

Data to be computed

• Measured data: Specified with channel numbers (example: 050).
• Computed data: Specified with computation channel numbers.
• Constants: 30 constants for the stand-alone type and 60 constants for the expanded

type can be set. Each constant is expressed by a number from K01 to K60 (K01 to

K30 for the stand-alone type).

Priority of operators

The priority of operators in a computing equation is as follows. The operators are placed

in order from the highest priority.

Type

Operators

Function

ABS(), SQR(), LOG(), LN(), EXP(), MAX(), MIN(), P-P(), SUM(), AVE(),
PRE(), HOLD():, RESET():, PSUM()

Power

**

Signs, logical negation

+, -, NOT

Multiplication, division

*, /

Addition, subtraction

+, -

Greater/less relation

.GT., .LT., .GE., .LE.

Equal/not equal relation .EQ., .NE.
Logical product

AND

Logical sum, exclusive

OR, XOR

Range when computing

When the value exceeds ±10308 during the computation, a computation error (overflow)

occurs.

Units in computing equations

In computations, measured data are handled as numbers without units. For example, if

the measured data of channel 001 is 20 mV and the measured data of channel 002 is 20 V,

the computed result of 001+002 becomes 40.

Limitations in computing equations

Multiple operators can be used in one computing equation. But, there are following

limitations.

• The number of characters that can be used : 40 characters

• The total amount of channel numbers and constants: 16 (Computation error occurs

when 16 is exceeded, and the computed result becomes +OVER or –OVER)

• Computation channel numbers: Computation channel numbers less than the current

computation channel number can be used as variables within the computing equation.

Example: A02=001+A01 Computation channel numbers greater than or equal to A03

can not be used in this computation.

• Statistical operators (TLOG. or CLOG.) can only be used once in one computing

equation.