Process of overflow data during TLOG or CLOG (TLOG CH OVER)

Used to select the process method for overflow data obtained during TLOG or CLOG.
• ERROR : Overflow data processed as a computation error.
• SKIP : Overflow data ignored and computation is continued.
• LIMIT : Overflow data treated as the next data for computation.
Measurement channels to which linear scaling is applied: Specified scaling upper-/lower-
limit.
Measurement channels to which no linear scaling is applied: Upper-/lower-limit of the
measuring range.
Computation channels: Specified LEFT/RIGHT value.

Processing of the results of TLOG.PSUM computation for pulse input

• OVER :

If the result of a separate calculation of TLOG.PSUM (XXXX) exceeds
99999999, the instrument goes into an overflow.

• ROTATE : If the result of a separate calculation of TLOG.PSUM (XXXX) exceeds

99999999, the instrument resets the value subsequent to 99999999 to 0

and continues computing. For example, if the measured data subsequent

to 99999999 which has been reached is 3, the computed result becomes 2

by counting 0, 1 and 2.

The process is effective only when a pulse input module is installed. This setting cannot

be made for a DR standalone type

Temperature Unit (for /D2.. deg F Display only)

If your DARWIN hardware is equipped with the /D2 option it is possible to change the
Temperature Unit (C/F) for TC and RTD.
Make sure to activate the Temperature option in the System Configuration setting screen,
by checking the Temperature Unit Switch (/D2).
Note, that as soon as you select Celsius or Fahrenheit the SET (Regular) Settings will all

be initialized. Therefore, make sure to switch the Temperature Unit before changing any

SET (Regular) settings.

SCSI ID Number (for DC100 only)

Set the ID so that it does not overlap with other SCSI devices connected to the DC100.

The default value is seven.

4.24 Setting (TLOG) Math Setup, Temperature Unit, and (DC100 only) SCSI ID Number