Calibration

Explanation

Perform the calibration in the following sequence:

Universal Input Module

20mV ZERO > 20mV SPAN > 60mV ZERO > 60mV SPAN > 200mV ZERO

> 200mV SPAN > 2V ZERO > 2V SPAN > 6V ZERO > 6V SPAN

> 20V ZERO > 20V SPAN > 50V ZERO > 50V SPAN > Pt:1mA ZERO

> Pt:1mA SPAN > Pt:2mA ZERO > Pt:2mA SPAN > Pt:1mA-H SPAN

> Pt:2mA-H SPAN > Cu:2mA ZERO > Cu:2mA SPAN

DCV/TC/DI Input Module

Same sequence as the Universal Input Module except for the RTD (resistance

temperature detector).

mA Input Module

20mA ZERO > 20mA SPAN

Strain Input Module

2k ZERO > 2k SPAN > 20k ZERO > 20k SPAN > 200k ZERO > 200k SPAN

Digital Module

60mV ZERO > 60mV SPAN > 6V ZERO > 6V SPAN

Adjustable range/Error conditions/Display

The adjustable range is from -32768 to 32767.

However, 16384, –16384, 16385, and –16383 are excluded.

Note

The adjustable range is as described above. However, if the following conditions are not met,

the module is considered erroneous.

• ZERO calibration value: –3277 to 3277. The ideal value is 0.

• SPAN calibration value: 29491 to –29491. The ideal value is –32768.

19000 to 21000 for the strain input module. The ideal value is 20000.

(21000)

–29491

(19000)

29491

(20000)

–32768

ZERO calibration value

SPAN calibration value

–3277

0

3277

Values in parenthesis are for strain input modules.

8.3 Manual Calibration