3 manual calibration, 3 manual calibration -4, Operating procedures – Yokogawa DAQWORX User Manual

Page 177

Advertising
background image

8-4

IM WX102-01E

8.3

Manual Calibration

Operating Procedures

1

Click the connect button on the calibration screen. The DA100, DC100, or DR

system that is connected is displayed.

2

Clicking the module to be calibrated displays a calibration dialog box.

3

Click the Manual tab to display the manual calibration screen.

Zero current value

Zero calibration value

Span current value

Span calibration value

Displays input range

Displays input range

Start button

Range

Select all

ranges

Manual tab

4

Clicking the range to be calibrated shows the selected range in red. To calibrate

all ranges click All.

5

Set the calibration values for Zero and Span, individually. Clicking the Zero or

Span display area shows the setting range.

6

Clicking the Start button displays a message for writing the calibration value.

7

Clicking the OK button displays a final message for fixing the calibration value.

8

Clicking the OK button displays a message notifying that the calibration value has

been fixed.

9

Clicking the OK button terminates the calibration. To calibrate another module,

click the Close button and repeat from step 2. OK is displayed for the modules

that have completed the calibration.