Calibration

5

Clicking the Start button displays the connection type of the terminal and the input

value in an illustration and a message for confirming the connection and input.

Auto tab

Message confirming

the wiring and input

value

Input voltage

6

Wire the terminals of the DA100, DC100, and DR as shown in the displayed

illustration and apply the specified voltage to the specified terminal.

7

If you had selected multiple ranges to be calibrated, the wiring and the input value

for the next range is shown in an illustration, just like in step 5, along with the

confirming message when the OK button is clicked.

8

After confirming the wiring and the input for all ranges according to steps 5 and

6, a final confirming message is displayed to fix the calibration value. Clicking the

OK button displays a message notifying that the calibration value has been fixed.

9

Clicking the OK button terminates the calibration. To calibrate another module,

click the Close button and repeat from step 2. OK is displayed for the modules

that have completed the calibration.

8.2 Auto Calibration