Yokogawa DAQWORX User Manual

Page 59

Advertising
background image

4-12

IM WX102-01E

RL : Rate-of-change lower-limit alarm. An alarm occurs when the measured value

changes in the decreasing direction within a certain time (rate-of-change alarm

interval) and exceeds the lower-limit alarm setpoint.

dH : An alarm occurs when the difference between two channels exceeds the upper-

limit alarm setpoint. This only applies to a channel for which interchannel

differential computation is selected, and can be set as a type of alarm.

dL : An alarm occurs when the difference between two channels exceeds the lower-

limit alarm setpoint. This only applies to a channel for which interchannel

differential computation is selected, and can be set as a type of alarm.

See also Section 4.20, “Setting Alarm Interval, Hysteresis, and Alarm Display Hold.”

Only H (upper-limit alarm) and L (lower-limit alarm) are available for computation

channels.

Note

• Alarms can’t be set if the input type is SKIP, or if the Math channel mode is OFF.

• If you make below changes, channel alarms will automatically be switched OFF:

• Input type.

• Measurement range.

• Reference channels of inter-channel computations.

• RRJC reference channels, reference channel inputs, or thermocouples.

• Setting Math channels ON, OFF, Math equations

Setting alarm values

Set one alarm value for each alarm number within the following range, depending on the

type of input. The default value is 0 (zero).

H and L

• VOLT, TC, RTD, DI, DELTA, AC, STRAIN, mA, and PULSE: Within the measuring

range in the related channel. DI can be set to either 0 or 1.

• SCALE: Within a linear scaling value.
• Math channel: Within the range specified by the Math function.

RH and RL

• VOLT, TC, RTD, DI, and DELTA, AC, STRAIN, mA, and PULSE: Within 1 to [maximum

measuring range minus minimum measuring range] in the related channel. For

example, set 30000 (without using a decimal point) if the maximum value exceeds

30000. DI can be set to only 1.

• SCALE: Set 1 to 30000 without using a decimal point.

dH and dL

• When the reference channel uses VOLT: Within the measuring range.

• When the reference channel uses other than VOLT: [maximum measuring range

minus minimum measuring range] to - [maximum measuring range minus minimum

measuring range]

4.4 Alarm Settings

Advertising
This manual is related to the following products: