Recording-related Settings (DR Only)

In this Section the following recording-related setting for DR:

• Recording zones

• Partially expanded recording

• Interpolation

• Trend recording

Digital printout ON/OFF

Manual printout ON/OFF

Scale printout ON/OFF

List printout ON/OFF

Setting Procedure

1

To open the Channel Settings related setting screen, click the Channels tab, or

select Settings > SET (Regular) Settings > Channel Settings from the menubar.

For the Math channel related setting screen click the Math tab, or select Settings

> SET (Regular) Settings > Math Functions from the menubar.

Left recording zone

Right recording zone

List print ON/OFF

Scale value print ON/OFF

Manual print ON/OFF

Digital print ON/OFF

Trend recording ON/OFF

Boundary value

Interpolation ON/OFF

Partially expanded

recording settings

2

To specify the recording zone, set the left and right position. The distance between

the two sides should be more than 55 mm.

3

Set partially expanded recording ON or OFF, and set the Position and Boundary.

Position and Boundary can’t be set if partially expanded recording is OFF.

4

Set Interpolation ON or OFF. If set to ON, interpolation will be carried out

according to the priority order of recording colors.

6

Set Trend Print ON or OFF. Only channels set ON will get a trend printout.

7

Set Manual Print ON or OFF. Only channels set ON will get a manual printout.

8

Set Scale Print to ON1/ON2/ON3 to print Scale values during trend recording, or

OFF.

9

Set List Print ON or OFF. The contents of channels set to ON will be included in

the List printout.

Setting Recording Zones (ZONE)

The set left and right position of the zone correspond to the left and right span set at the

SPAN menu (recording span). Zones can be set in 1mm steps and cannot exceed the

recording range (i.e. 250mm). The left position setting ranges from 0 to 245mm. Right

position setting ranges from 5 to 250mm. The minimum width of a zone is 5mm. No

decimal points. Default settings: 0mm for the left position, 250mm for the right position.

If the instrument is equipped with the optional Math function, or if it has a pulse module
or a FDD (DR232/DR242) attached, this setting can also be made for computation
channels A01 to A60.