Resistance:static:recall, Resistance:static:low:avalue/bvalue – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 105

COMMAND DETAILS
105
Return Parameter <NR2> [MAX|MIN]
Slew rate
1 Unit =1 amp/uS
Returns the slew rate.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:RES:RISE? MAX
0.8000
Returns the maximum
value for the rising slew
rate (0.8 A/uS).
:RESistance:STATic:RECall
Channel Specific
Command
Description
Sets or queries whether A Value or B Value is the
currently active value in CR static mode.
Syntax
:RESistance:STATic:RECall {A|0|B|1}
Parameter
A/0
A
B/1
B
Example
:RES:STAT:REC 1
Makes B Value the active
value.
Query Syntax
:RES:STATic:RECall?
Return Parameter <NR1>
Value
0
A
1
B
:RESistance:STATic:LOW:AVALue/BVALue
Channel Specific
Command
Description
Sets the low range A/B Value for constant
resistance static mode.
Syntax
:RESistance:STATic:LOW:AVALue/BVALue
<NRf+>[OHM]
Parameter
NRf+[OHM]
Resistance
AVALue 10
Sets A Value to 10 ohms.
(Low range only)