Configure:reset, Configure:group:units – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 49

COMMAND DETAILS
49
Example
:CONF:RESP 0
Response set to normal.
Query Syntax
: CONFigure:RESPonse?
Return Parameter <NR1>
Response
0
Normal
1
Fast
Query Example
:CONF:RESP?
1
Response is Fast.
:CONFigure:RESEt
Channel Specific
Command
Description
Recalls the original factory default settings.
Syntax
:CONFigure:RESEt
Example
:CONF:RESE
:CONFigure:GROup:UNITs
Channel Specific
Command
Description
Sets or queries the number of single channel load
modules (PEL-2040 or PEL-2041) that can be used
in the parallel mode.
Syntax
CONFigure:GROup:UNITs <NRf>|MIN|MAX
Parameter
<NRf>
Number of units
MIN
Sets to the minimum
number
MAX
Sets to the maximum
number
Example
CONF:GRO:UNIT 2
Sets the parallel mode to 2
units.
Query Syntax
CONFigure:GROup:UNITs? [MIN|MAX]
Return Parameter <NR1>
Returns the number of
units
MAX/MIN
Returns the allowable
maximum and minimum.