Current:dynamic:t1/t2 – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 64

PEL-2000 Programming Manual
64
Return Parameter <NR2> [MAX|MIN]
Slew rate
1 Unit=1 amp/uS
Returns the slew rate.
MAX/MIN
Returns the allowable
maximum or minimum.
Query Example
:CURR:DYN:FALL? MIN
0.0003
Shows the minimum
allowable value for the
falling slew rate as 0.0003
A/uS for the specific
channel and range.
:CURRent:DYNamic:T1/T2
Channel Specific
Command
Description
Sets the timers T1 or T2 for CC dynamic mode for
the specific channel and range.
Syntax
:CURRent:DYNamic:T1/T2 <NRf+>[S|ms]
Parameter
<NRf+>[S]
Time
T1/T2 0.1S
Sets the T1/T2 time to 0.1
seconds.
T1/T2 1
Sets T1/T2 to 1 second.
T1/T2 MIN
Sets the T1/T2 to the
minimum value.
T1/T2 MAX
Sets the T1/T2 time to the
maximum time
Example
:CURR:DYNA:T1 .1S
Sets the T1 time to 100
milliseconds for the
specific channel.
Query Syntax
: CURRent:DYNamic:T1/T2? [MIN|MAX]
Return Parameter <NR2> [MAX|MIN]
Time
1 Unit=1 second
Returns T1/T2 time.
MAX/MIN
Returns the allowable
maximum and minimum.