Current:dynamic:l1/l2 – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 62

PEL-2000 Programming Manual
62
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.
Example
:CURR:STAT:HIGH:RISE
1.1
Sets the rising slew rate
to 1.1A/uS.
Query Syntax
:CURRent:STATic:HIGH:RISE/FALL? [MIN|MAX]
Return Parameter <NR2> [MAX|MIN]
Slew rate
1 Unit=1 amp/uS
Returns the slew rate.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:CURR:STAT:HIGH:RISE?
MAX
0.8000
For high range CC mode,
the maximum value for
the rising slew rate is
0.8000 A/uS for the
specific channel.
:CURRent:DYNamic:L1/L2
Channel Specific
Command
Description
Sets the current levels (Level 1 & 2) for CC
dynamic mode. The command is range dependant.
If the current range is Low, then the settings will
only apply to low range.
Syntax
:CURRent:DYNamic:L1/L2 <NRf+>[A]
Parameter
NRf+[A]
Current
L1 1
Sets L1 to 1 Amp.
L2 2
Sets L2 to 2 Amps.
L2 2A
Sets L2 to 2 Amps.
L1/L2 MIN
Sets L1 or L2 to the
minimum level for the
specific channel.