Sequence:voltage:range – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 154
Advertising

PEL-2000 Programming Manual
154
Parameter
OFF/0
Off
ON/1
On
Example
:SEQ:END:LOAD 0
On End Load = Off.
Query Syntax
:SEQuence:END:LOAD?
Return Parameter <NR1>
On End Load
0
Off
1
On
Query Example
:SEQ:END:LOAD?
1
On End Load is On.
:SEQuence:VOLTage:RANGe
Channel Specific
Command
Description
Sets the sequence CC voltage range.
Syntax
:SEQuence:VOLTage:RANGe <NRf>[V]|L|H
Parameter
<NRf>[ V] , L, H
Range
16
Low range*
80V
High range*
L
Low range
H
High range
*Load module dependent, PEL-2020 shown.
Example
:SEQ:VOLT:RANG L
Sets the range to Low for
the channel.
Query Syntax
:SEQuence:VOLTage:RANGe?
Return Parameter <NR2>
Range
16
Low PEL-2020,2030,2040
125
Low PEL-2041
80
High PEL-2020,2030,2040
500
High PEL-2041