Ocp:channel:end, Ocp:channel:step:current – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 89

COMMAND DETAILS
89
:OCP:CHANnel:END
Channel Specific
Command
Description
Sets the ending current value for the test. The
value must be higher than the DUT OCP value.
Syntax
:OCP:CHANnel:END {<NRf>[ A ] | MIN | MAX}
Parameter
<NRF>[A]
The current value in Amps.
MAX
The maximum current value.
MIN
The minimum current value.
Example
:OCP:CHAN:END MIN
Set the ending current to
the minimum.
Query Syntax
:OCP:CHANnel:END?
Return Parameter <NR2>
Returns the ending current in Amps.
Query Example
:OCP:CHAN:END?
10.0
Returns the ending
current.
:OCP:CHANnel:STEP:CURRent
Channel Specific
Command
Description
Sets the current step resolution for the OCP Test
Automation.
Syntax
:OCP:CHANnel:STEP:CURRent {<NRf>[ A ] | MIN |
MAX}
Parameter
<NRF>[A]
The current value in Amps.
MAX
The maximum current value.
MIN
The minimum current value.
Example
:OCP:CHAN:STEP
:CURRent MIN
Set the step resolution to
the minimum value.
Query Syntax
:OCP:CHANnel:STEP:CURRent?
Return Parameter <NR2>
Returns the current step resolution in
Amps.
Query Example
:OCP:CHAN:STEP:CURR?
0.5
Returns the step
resolution.