133
MAX
Sets the current limit to
the maximum value for the
channel
Example
:VOLT:CURR 1A
Sets the current limit to 1
Amp for the specific
channel.
:VOLT:CURR MAX
Sets the current limit to
the maximum limit for
the specific channel.
Query Syntax
:VOLTage:CURRent? [MAX|MIN]
Return Parameter <NR2> 1 unit = 1 amp
Current limit
1
Returns the current limit of
the specific channel.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:VOLT:CURR?
5
The current limit is 5
amps for the specific
channel.
:VOLTage:MODE
Channel Specific
Command
Description
Set the constant voltage response time for the
specific channel.
Syntax
:VOLTage:MODE {SLOW|0|FAST|1}
Parameter
SLOW/0
Slow response time
FAST/1
Fast response time
Example
:VOLT:MODE SLOW
Sets the response time to
SLOW for the specific
channel.
:VOLT:MODE 1
Sets the response time to
FAST for the specific
channel.
Query Syntax
:VOLTage:MODE? <NR1>