Sequence:loop:start, Sequence:channel:time – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 155

COMMAND DETAILS
155
Query Example
:SEQ:VOLT:RANG?
500
Returns the voltage
range. In this case high
for the PEL-2041.
:SEQuence:LOOP:STARt
Channel Specific
Command
Description
Determines from which point to start repeating
(looping) the sequence from when using the
SEQuence:REPeat command.
Syntax
:SEQuence:LOOP:STARt <NR1>
Parameter
<NR1>
Start loop from
1~last point
1st~ last point.
Example
:SEQ:LOOP:STAR 2
Loop from point 2.
Query Syntax
:SEQuence:LOOP:STARt?
Return Parameter <NR1>
Point
1~last point
Returns the point that the
loop will start from.
Query Example
:SEQ:LOOP:STAR?
2
The loop starts at point
2.
:SEQuence:CHANnel:TIME
Channel Specific
Command
Description
Sets which channel duration time the specific
channel will use.
Syntax
:SEQuence:CHANnel:TIME <NR1>
Parameter
<NR1>1~max channels
Duration Time Settings
1
Use channel 1
2 etc
Use channel 2 etc
Example
:SEQ:CHAN:TIME 3
Set the specific channel
to use the channel
duration time of channel
3.