L1/L2 MIN
Sets L1 or L2 to the
minimum level for the
specific channel.
L1/L2 MAX
Sets L1 or L2 to the
maximum level for the
specific channel.
Example
:RES:DYN:HIGH:L1 10
In high range CR
dynamic mode, Set L1
(level 1) to 10 ohms.
Query Syntax
:RESistance:DYNamic:HIGH:L1/L2? [MIN|MAX]
Return Parameter <NR2> [MAX|MIN]
Resistance
1 unit= 1 ohm
Returns the resistance of
Level 1/ 2 (L1/L2).
Query Example
:RES:DYN:HIGH:L2? MAX
15000.0
Returns the maximum
resistance allowed for
the channel. (PEL-2020)
:RESistance:DYNamic:HIGH:RISE/FALL
Channel Specific
Command
Description
Sets the high range rising/falling slew rate for CR
dynamic mode for the specific channel.
Syntax
:RESistance:DYNamic:HIGH:RISE/FALL
<NRf+>[A/uS]
Parameter
<NRf+>[A/uS]
Slew Rate
RISE/FALL 0.8A/uS
Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL 1
Sets the rising/falling slew
rate to 1A/uS
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.
Example
:RES:DYN:HIGH:RISE 1.1
Sets the rising slew rate
to 1.1A/uS.