Resistance:dynamic:high:l1/l2 – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 111

COMMAND DETAILS
111
Parameter
<NRf+>[S]
Time T1/T2
T1/T2 0.1S
Sets the T1/T2 time to 0.1
seconds.
T1/T2 1
Sets T1/T2 to 1 second.
T1/T2 MIN
Sets the T1/T2 to the
minimum value.
T1/T2 MAX
Sets the T1/T2 time to the
maximum time
Example
:RES:DYNA:LOW:T1 10S
Sets the T1 time to 10
seconds for the specific
channel.
Query Syntax
:RESistance:DYNamic:T1/T2? [MIN|MAX]
Return Parameter <NR2> [MAX|MIN]
Time T1/T2
1 Unit=1 second
Returns T1/T2 time.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:RES:DYN:LOW:T1?
2.5
:CURR:DYN:LOW:T1?
MIN
0.000025
Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.
:RESistance:DYNamic:HIGH:L1/L2
Channel Specific
Command
Description
Sets the high range resistance levels (Level 1 & 2)
for CR dynamic mode.
Syntax
:RESistance:DYNamic:HIGH:L1/L2 <NRf+>[OHM]
Parameter
NRf+[OHM]
Resistance
L1 10
Sets L1 to 10 ohms. (high
range only)
L2 20OHM
Sets L2 to 20 ohms. (high
range only)