Current:dynamic:rise/fall – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 63

COMMAND DETAILS
63
L1/L2 MAX
Sets L1 or L2 to the
maximum Level for the
specific channel.
Example
:CURR:DYN:L1 10
In CC dynamic mode,
Set L1 (level 1) to 10
amps.
Query Syntax
:CURRent:DYNamic:L1/L2? [MIN|MAX]
Return Parameter <NR2> [MAX|MIN]
Current
MAX/MIN
Returns the allowable
maximum and minimum.
1 unit= 1 amp
Returns the current of
L1/L2, or the maximum or
minimum current allowed.
Query Example
:CURR:DYN:L2?
2.0400
Returns current for the
specific channel.
:CURRent:DYNamic:RISE/FALL
Channel Specific
Command
Description
Sets the rising/falling slew rate for CC dynamic
mode for the specific channel and range.
Syntax
:CURRent:DYNamic:RISE/FALL <NRf+>[A/uS]
Parameter
<NRf+>[A/uS]
Slew Rate
RISE/FALL 0.8A/uS
Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL 1
Sets the rising/falling slew
rate to 1A/uS
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.
Example
:CURR:DYNA:RISE 1.1
Sets the rising slew rate
to 1.1A/uS.
Query Syntax
: CURRent:DYNamic:RISE/FALL? [MIN|MAX]